MRI-1 Ventilator

Référence du produit MRI-1

The MRI-1 Ventilator is a small animal ventilator designed for use in MRI and other high magnetic field environments. The system comprises a microprocessor-based control unit and a set of remote, pneumatically operated, non-metallic valves. The MRI-1 can operate by itself or can be controlled and/or monitored by a computer.
The MRI-1 operates on the flow-time principle: an inspiratory airflow is established and gated into the animal for a known time, thus producing a known volume. This approach provides extraordinary flexibility — a wide range of volumes, breaths-per-minute, and I/E ratios are possible with no hardware changes, and using just three front-panel controls.
Respiratory airflow is provided by an internal airpump. Alternatively, an external pressurized gas source can be attached (oxygen or anesthetic gasses). The remote high-speed, miniature pneumatic valves do the actual airflow switching. By locating these valves close to the animal (including inside the magnet bore if necessary), dead-space and tubing compliance are minimized.

Description

MRI-1 Ventilator

  • MRI compatible ventilator
  • Non-magnetic pneumatic valves (XS, S, & M sized valves available depending on weight of animal)
  • Dependable all-electronic ventilator
  • Wide tidal volume and rate range
  • Mice to Monkeys (up to 10kg with appropriate valve & accessories)
  • Safe with oxygen and anesthesia
  • Direct readout of volume and rate
  • Easy to set up

Technical Specifications ;

  • Respiratory rate range:  5 – 200 breaths/min
  • Inspiratory flow range:  60 – 1000ml/min
  • Inspiration/Expiration (I/E ratio) range:  20 – 80%
  • Tidal volume range:  0.1 – 30ml
  • Manual control functions:  HOLD (suspends inspiration), MAN (manual inflation)
  • LCD panel:  4 lines x 20 chars, backlit
  • Indicator lights:  INSPIRATION, LOW PRES WARNING, ERROR
  • Internal airpump capacity:  4 lpm
  • Pneumatic valve switching speed (air actuation):  < 75mS
  • Pneumatic valve switching speed (Helium actuation):  < 55mS
  • Pneumatic valve actuation pressure:  3 – 5 bar
  • Pneumatic valve flow orifices:  0.080 in. (2.0mm)
  • Pneumatic valve materials:  PTFE body, non-metallic fasteners, one non-ferrous coil spring
  • Dimensions, main unit:  9W x 5.5H x 9D in. (23 x 14 x 23cm)
  • Dimensions, pneumatic valve (each, of two):  1.0 dia x 1.8 long, in. (47 x 25mm)
  • Port connections, respiratory and pressure functions:   0.125″ (3mm) ID Tygon tubing
  • SYNC OUT connector:  BNC
  • Data Port connector:  DB-25S
  • Power requirements:  120/240V (switchable), 100VA


Produits apparentés

  • AD-5000 Système d’anesthésie Numérique

    NOUVEAU

    Le AD-5000 Système d’Anesthésie Numérique comprend un écran de commande tactile unique et clair offrant à l’opérateur une gamme complète de fonctionnalités pratiques, y compris un contrôle de débit précis et simultané de 0,5 à 2,5 l / min pour deux masques et une chambre d’induction. Le circuit de débit exclusif permet à l’utilisateur de choisir n’importe quelle combinaison de sorties tandis que le contrôle de débit numérique s’ajuste automatiquement pour donner les débits corrects. La chambre d’induction et les circuits pour les masques comprennent une minuterie pour offrir une surveillance précise et éviter une surexposition au sujet.

    Ajouter à votre panier
  • SAR-1000 Ventilateur volumétrique et barométrique avancé

    Le SAR-1000 Ventilateur volumétrique et barométrique fonctionne comme un ventilateur volumétrique ou barométrique pour rats, lapins, cobayes et en option, avec une version rajoutée pour les souris. Le mode VOLUME livre un volume courant fixe à chaque respiration et le mode PRESSION provoque l’arrêt de l’inflation quand il a atteint une pression d’air préréglée. Dans les deux modes, l’inflation des poumons est maintenu à la fin de l’inspiration jusqu’au moment où le temps de pourcentage d’inspiration (%Insp) est écoulé.

    Ajouter à votre panier
  • EZ-SA800 Système d’anesthésie de traitement individuel

    Le EZ-SA800 Système d’anesthésie de traitement individuel permet d’être compatible pour des animaux allant des souriceaux nouveau-né aux lapins et est conçu pour anesthésier un seul animal à la fois. Ces deux valves permettent à l’ anesthésiant d’aller directement vers la chambre d’induction et/ou vers le respirateur.

    Ajouter à votre panier